MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/19/2021
Event Type  malfunction  
Event Description
It was reported that upon power up of the cardiosave intra-aortic balloon pump (iabp) there was a technical error #6.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Communication/interviews: a getinge field service engineer (fse) was dispatched to investigate.The fse conversed with the biomedical technician over the phone and instructed him to enter into service mode as to calibrate the touch screen.Soon thereafter, the biomedical technician was prompted to calibrate the touchscreen.After successfully completing calibration, the biomedical technician powered the unit back on into clinical mode and the technical error 6 no longer alarmed.The biomedical technician left the pump running for an hour and confirmed no failure occurred.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13001327
MDR Text Key285158773
Report Number2249723-2021-02854
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-