MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9734252, serial/lot #: unknown. A medtronic representative went to the site to test the equipment. The system then passed the system checkout and was found to be fully functional. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a cranial biopsy procedure. During a procedure, navigation was lost while performing a biopsy. After patient registration and swapping to the sterile reference frame, the "navigation was out. " the surgeon thought the reference frame may have moved during the draping process or when changing from the non-sterile reference frame to the sterile reference frame. The articulating arm was checked. When fully tightened , the arm would move. The articulating arm would be replaced. The procedure was completed without the use of imaging or navigation. The reported issue resulted in a procedure delay of an hour or longer. The impact to the patient was not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION¿ S8 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13003543
MDR Text Key288608606
Report Number1723170-2021-02886
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 12/14/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/03/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-