• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problems Break (1069); Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  malfunction  
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that a rt380 adult dual-heated evaqua2 breathing circuit was found cracked during use.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).The complaint rt380 adult dual-heated evaqua2 breathing circuit is expected to but has not yet been received at fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event Description
A healthcare facility in spain reported via a fisher & paykel healthcare (f&p) field representative that a rt380 adult dual-heated evaqua2 breathing circuit was found cracked during use.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).Corrected data: section d9 device available for evaluation? no.Section h3 device evaluated by manufacturer? other method: the complaint rt380 adult dual heated evaqua2 breathing circuit was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photograph provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the provided photograph revealed that the evaqua 2 tube was degraded.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.However, based on previous investigations of similar complaints, the reported event was likely due to the tube being in contact with chemicals, either nebuized drugs or cleaning solutions.All rt380 adult evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject adult breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt380 adult evaqua2 breathing circuit state the following: - "do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "set appropriate ventilator alarms.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key13006728
MDR Text Key284005956
Report Number9611451-2021-01422
Device Sequence Number1
Product Code BZE
UDI-Device Identifier09420012429728
UDI-Public(01)09420012429728(10)2100826200(11)190717
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K122432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot Number2100826200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-