The company representative confirmed and replicated the reported event.The outbreak printed circuit board (pcb was subsequently replaced to resolve the issue.The system was tested and found to meet product specifications.The manufacturing device history record (dhr) was reviewed.The dhr was completed and reviewed by qa to ensure that the product was manufactured in compliance with the device master record.Based on qa assessment, the product met specifications.The root cause of the reported event is attributed to nonconforming breakout pcb.Based on a low occurrence rate, it is determined that no further actions are necessary at this time.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|