Model Number 9733560
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9 733467, serial/lot #: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess.) it was reported that the system was flickering between green and red status during navigation.The facility noted that the reported issue occurred exclusively with instruments that used instrument trackers and navigated blades did not have the reported issue.This issue occurred intraoperatively and caused no surgical delay.There was no impact on patient outcome.Additional information received indicated the troubleshooting performed was unknown.The reported issue sounded like metal interference/emitter positioning.The manufacturer representative attended another case with the surgeon and was unable to replicate the behavior.The manufacturer representative was going to perform an in-service to ensure the site was using the system correctly.The issue was considered to have resolved during the call.
 
Manufacturer Narrative
H3) the logs for the navigation system were returned to the manufacturer for evaluation.Analysis found that logs contained insufficient information to determine the relationship of the software to the reported issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® ENT NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13012148
MDR Text Key287117868
Report Number1723170-2021-02898
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169504394
UDI-Public00643169504394
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9733560
Device Catalogue Number9733560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age25 YR
Patient SexFemale
Patient Weight68 KG
-
-