Model Number 3660 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problems
Failure of Implant (1924); Inadequate Pain Relief (2388)
|
Event Date 11/01/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of the event is estimated.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|
|
Event Description
|
It was reported that the implantable pulse generator was inoperable.Patient lost therapy.Patient underwent an unrelated surgery and did not place the device in surgery mode.A surgical intervention is anticipated in the near future.No additional information is available at this time.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure proper packaging and sterility.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Additional information was received indicating that the patient's ipg was recovered by an abbott representative and therapy was restored to the patient.
|
|
Manufacturer Narrative
|
The "report type" has been changed to "malfunction".Section b1: uncheck "adverse event".Section b2: uncheck "other serious (important medical events)".Section h7: remedial action unchecked and removed.Section h9: fsca has been removed due to the ipg being recovered.An inoperable ipg was reported to abbott.The system was not set to surgery mode while the patient underwent an unrelated surgery where electro-cautery was used.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with unintentional damage during the unrelated surgery.The system was recovered through abbott intervention.The ipg was not replaced to reestablish effective therapy.
|
|