• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW¿; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW¿; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE Back to Search Results
Catalog Number 688508/JPA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that during the hemodynamic monitoring of a patient, the pressure monitoring set was found to be leaking.No patient injury to report.
 
Manufacturer Narrative
The suspect device has been returned for evaluation.The device was examined visually, and functional testing was performed.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDRAW¿
Type of Device
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, 76892 6
SN  768926
Manufacturer (Section G)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, 76892 6
SN   768926
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
MDR Report Key13014836
MDR Text Key284985905
Report Number8020616-2021-00096
Device Sequence Number1
Product Code DXN
UDI-Device Identifier04589731124477
UDI-Public4589731124477
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/09/2023
Device Catalogue Number688508/JPA
Device Lot NumberC2105580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/31/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received12/31/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-