Model Number URF-P6
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to the olympus local service department. A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event Description
Olympus medical systems corp. (omsc) was informed that during the incoming inspection for repair at the olympus local service department, it was found that the internal metal wire had been broken and exposed from the bending section of subject device since the bending tube sheath was broken. Other detailed information was not provided. There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO FIBERSCOPE
Type of DeviceURETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13016181
MDR Text Key282350462
Report Number8010047-2021-16237
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/28/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-