Medtronic received a report that the pipeline failed to open, experienced resistance during retrieval, and the pushwire broke.
The patient was undergoing treatment for an unruptured, saccular aneurysm located in the paraclinoid segment of the left internal carotid artery.
The patient's vessel tortuosity was moderate.
The access vessel was the femoral artery, which was 2.
0mm in diameter.
The landing zone was 4.
8mm distal and 5.
1mm proximal.
It was reported that the pipeline did not open distally even after massaging the device.
After trying for several minutes and not su cceeding in opening the device, the doctor decided to remove it.
When trying to retrieve the device it became stuck in the proximal hub of the microcatheter.
When attempting to retrieve the device, the distal shaft of the delivery system broke.
There was no damage to the catheter.
A continuous flush had been administered.
no further hospitalization or medical surgical interventions were needed, and the patient did not experience any injury or complications.
The devices were prepared and flushed according to the instructions for use (ifu).
Ancillary devices include a sofia catheter and marksman microcatheter.
|