BAXTER HEALTHCARE CORPORATION I.V. ADMINISTRATION SETS WITH CONTROL-A-FLO REGULATOR; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number EMC5905P |
Device Problem
Filling Problem (1233)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the flow rate regulator of a flow regulator set could not adjust the drip rate.The reporter stated that the regulator dial was able to be rotated, but did not change the flow rate.The flow rate was faster than expected.This occurred prior to patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the actual device was received for evaluation enclosed in its original packaging.Visual inspection did not reveal any issues however, the sample was received in a condition that could not could not be functionally tested.Therefore, the reported condition could not be verified on the actual sample.Meanwhile, retention samples were visually inspected with no issues noted.The retention samples underwent gravity and functional testing and were found to be conforming to product specifications.The reported condition was not verified by the evaluation of the retention samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|