Model Number 191126
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2021
Event Type  malfunction  
Event Description
A user facility biomedical technician (bmt) reported that a 2008t machine experienced a bic pump no eos message during treatment. Additionally, a blown c1 capacitor with scorch marks was discovered on the motherboard during the machine evaluation, after patient treatment was complete. Additional information was requested, however to date has not been provided.
 
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.  .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T HD SYS. CDX W/BIBAG BLUE STAR
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key13021086
MDR Text Key282359265
Report Number2937457-2021-02461
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number191126
Device Catalogue Number191126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received01/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-