BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Lot Number 0026571625
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 14dec2021.It was reported that crossing difficulties were encountered.The 70-75% stenosed target lesion was located in the severely calcified and severely tortuous distal left circumflex artery.2.25 x 28mm synergy xd drug eluting stent was advanced but failed to cross the target lesion.The device was removed and the procedure completed with another of the same device.There were no patient complications.However, device analysis revealed stent damage.
 
Manufacturer Narrative
The synergy xd mr ous 2.25 x 28mm stent delivery system was returned for analysis.A visual examination of the stent found distal struts lifted.The undamaged crimped stent od (outer diameter) was measured within maximum crimped stent profile measurement.The balloon body was reviewed; no issues were noted on the balloon cones.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the tip showed no signs of damage.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and visual examination of the inner lumen found no issues with the extrusion shaft.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13021757
MDR Text Key282395295
Report Number2134265-2021-15936
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/17/2022
Device Lot Number0026571625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-