On b)(6) 2021, information was received from a patient receiving dilaudid (unknown dose and concentration) via an implantable pump for spinal pain.
It was reported the patient had been going though withdrawals for 4-5 days.
On the patient's 8835 ptm, they were getting code 8532 (stopped pump may exceed tube set).
The patient stated they have gone to the emergency room two times due to withdrawals.
The patient stated they could not function, their skin was crawling, could not sleep and could not "do anything".
The patient thought they were having an anxiety attack and they even changed the patient's anxiety mediations.
The patient's last refill was (b)(6) 2021.
The patient has not had any mris recently.
The patient stated they called their hcp and they have him coming in next week, thursday, to replace the pump.
The patient then stated the pump alarmed about 4-5 days ago and has not alarmed for the last 1.
5 - 2 days that they could recall.
End of service was reviewed with the patient.
The patient stated their hcp told them they were still getting dilaudid even though the pump was not functioning.
Troubleshooting was not required.
The issue was not resolved as troubleshooting was not needed.
The patient was redirected to their hcp to further address the issue.
|