The reported issue was confirmed.The root cause of the reported issue was overfill.Also explained how overfill occurred and stated it was a frequent issue.The device did not meet specifications, and was influenced by the reported failure.The device was in use on a patient.The device was not returned for evaluation.Dhr is not required if the failure is not out-of-box.Based on the results of the investigation, no additional actions are needed.The instructions for use were found adequate and state the following: "1) fill the reservoir with sterile or distilled water only.2) four liters of water will be required to fill the reservoir at initial installation.3) add one vial of arctic sun temperature management system cleaning solution to the sterile or distilled water 4) from the patient therapy selection screen, press the either the normothermia or hypothermia button, under the new patient heading.5) from the hypothermia or normothermia therapy screen, press the fill reservoir button.6) the fill reservoir screen will appear.Follow the directions on screen." the device was not returned.
|