MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problem Imprecision (1307)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2021
Event Type  malfunction  
Event Description
Medtronic received information regarding a navigation system. It was reported that during an em transsphenoidal (using ent software) case the surgeon felt inaccurate by 1 cm when deep in the anatomy. The inaccuracy was seen with two instruments, and both instruments were accurate on the surface of the skin. The site uses a mix of trace and touch for registration. The rep stated that the area of the inaccuracy was within the green zone of accuracy. The surgeon proceeded with navigation accounting for the inaccuracy. There was no patient impact and less than an hour of delay.
 
Manufacturer Narrative
A manufacturer rep went to the site to test the system. The system passed the system checkout. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION¿ S8 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13026027
MDR Text Key289668171
Report Number1723170-2021-02912
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 03/04/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No