Model Number 25198 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/18/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the device led display could not be lit.No patient impact or consequences were reported.
|
|
Event Description
|
The customer reported the device led display could not be lit.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: the returned device was evaluated.Visual inspection found no external damage or defects.The unit powered on and off using battery power.The two digits in the middle and right are missing from the top and bottom display.Internal inspection found flux contamination around the system board display driver component.The component was cleaned with isopropyl alcohol and the device was able to display all the digits.A service history record review reveals that this unit was in the field for over three (3) years with no previous reported issues related to this reported event.
|
|