• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RAD-5 PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RAD-5 PULSE OXIMETER Back to Search Results
Model Number 25198
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2021
Event Type  malfunction  
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event Description
The customer reported the device led display could not be lit.No patient impact or consequences were reported.
 
Event Description
The customer reported the device led display could not be lit.No patient impact or consequences were reported.
 
Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.Visual inspection found no external damage or defects.The unit powered on and off using battery power.The two digits in the middle and right are missing from the top and bottom display.Internal inspection found flux contamination around the system board display driver component.The component was cleaned with isopropyl alcohol and the device was able to display all the digits.A service history record review reveals that this unit was in the field for over three (3) years with no previous reported issues related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-5 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key13029825
MDR Text Key284988641
Report Number3019388613-2021-00358
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997005517
UDI-Public00843997005517
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K033296
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25198
Device Catalogue Number9196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Was the Report Sent to FDA? No
Date Manufacturer Received01/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-