ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804275-18
Device Problems Difficult to Remove (1528); Failure to Advance (2524); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2021
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis. A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint. Additionally, a review of the complaint history identified no other incidents from this lot. The investigation determined the reported difficulties appear to be related to the operational context of the procedure. There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event Description
It was reported that the procedure was to treat a heavily calcified, heavily tortuous mid left anterior descending artery. A 2. 75x18mm xience skypoint stent delivery system (sds) failed to cross due to anatomy. Once the sds was removed to perform dilatation with an unspecified balloon dilatation catheter the distal tip appeared to be flared and resistance with anatomy was also noted during removal. Balloon dilatation was performed and another unspecified stent was used to successfully complete the procedure. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameXIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Type of DeviceDRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13043452
MDR Text Key282498678
Report Number2024168-2021-11867
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 12/20/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1804275-18
Device Catalogue Number1804275-18
Device Lot Number1050641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-