ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804225-15
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2021
Event Type  malfunction  
Event Description
It was reported this was a procedure to treat a moderately calcified, mildly tortuous and 100% stenosed right posterior atrio-ventricular artery. During advancement of a xience skypoint 2. 25x15 drug eluting stent (des), it was observed during advancement, the stent moved but remains tightly crimped. The device was removed and replaced with another xience stent. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no similar incidents from this lot. The investigation determined the reported difficulties appear to be related to circumstances of the procedure. There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameXIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Type of DeviceDRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13044355
MDR Text Key284982133
Report Number2024168-2021-11888
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 12/20/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1804225-15
Device Catalogue Number1804225-15
Device Lot Number1031041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-