Catalog Number 3301-1
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3012447612-2021-00472.
 
Event Description
It was reported four sequoia closure tops stripped during final tightening in surgery.There were no patient impacts.This is report one of four for this event.
 
Event Description
It was reported four sequoia closure tops stripped during final tightening in surgery.There were no patient impacts.This is report one of four for this event.
 
Manufacturer Narrative
Device evaluation: visual inspection revealed that the screw drives of each closure top are slightly stripped.There was no visible damage to the threads.Potential cause root cause was unable to be determined.This event could possibly be attributed to off-axis forces applied during closure top insertion, cross-threading, or using a stripped driver.Dhr review per dhr reviews, the parts were likely conforming when they left zimvie control.Device use this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSURE TOP
Type of Device
SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key13046340
MDR Text Key282550858
Report Number3012447612-2021-00469
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00889024334861
UDI-Public(01)00889024334861(10)AAJ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K131980
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3301-1
Device Lot NumberAAJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-