A steris service technician arrived onsite to inspect the table and found the table to be operating properly, no repairs were required.While onsite, the technician learned that the table had an arthrex hds well leg spar long assembly attached at the time of the reported event.The technician was able to duplicate the reported event with the attachment installed.The 3085 surgical table operator manual states (1-2), "warning instability hazard: possible patient or user injury, as well as table or accessory failure, may result from using steris table accessories for other than their stated purpose or from using, on steris tables, accessories manufactured and sold by other companies." steris is not the manufacturer or distributor of the arthrex hds well leg spar - long assembly.Steris notified the manufacturer, arthrex, of the reported event to internally investigate and evaluate for reportability in accordance with 21 cfr part 803.The technician counseled user facility personnel on the proper use and operation of the 3085 surgical table, specifically to not use the arthrex hds well leg spar long assembly with their table.No additional issues have been reported.
|