• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY. DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY. DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 0500318E
Device Problem Fluid Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2021
Event Type  malfunction  
Event Description
A user facility nurse reported that a dialyzer blood leak occurred ten minutes after the initiation of the patient¿s hemodialysis (hd) treatment on a fresenius 4008s machine. The blood leak was noted as being an external blood leak that was visually observed at the hansen connector. Blood leak strips were not used to test for the presence of blood. The patient¿s estimated blood loss (ebl) was not provided upon followup. There was no patient injury, adverse events, or medical intervention required as a result of this event. The patient was restarted on the same machine and treatment completed successfully with new supplies. The complaint device was reported to be unavailable to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.  .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOPTIFLUX 180NRE DIALYZER FINISHED ASSY.
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key13047109
MDR Text Key285481310
Report Number1713747-2021-00467
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K002761
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0500318E
Device Catalogue Number0500318E
Device Lot Number21LU07012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received01/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/20/2021 Patient Sequence Number: 1
Treatment
FRESENIUS 4008S MACHINE; FRESENIUS 4008S MACHINE
-
-