Model Number U228
Device Problem Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2019
Event Type  malfunction  
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted. Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv)/respiratory sensor that is related to a high impedance condition. This device was included in the minute ventilation sensor signal oversensing advisory population. As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
It was reported that this device had two stored signal artifact monitor (sam) events. Evaluation of the first event identified it was generated from a false positive detection of atrial fibrillation. Evaluation of the second event identified it was generated from a false positive detection of an external electromagnetic interference source. A boston scientific technical services consultant stated the rate response minute ventilation (mv) sensor and the respiratory rate trend (rrt) had been disabled. The consultant recommended bringing the patient into the clinic to turn mv on if desired for blended sensor rate adaptive pacing. No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVISIONIST X4 CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13049918
MDR Text Key282556074
Report Number2124215-2021-36700
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559471
UDI-Public00802526559471
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 12/20/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/03/2021
Device Model NumberU228
Device Catalogue NumberU228
Device Lot Number738919
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2019
Is the Device Single Use? Yes
Type of Device Usage Initial

-
-