• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA PRO XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA PRO XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S40PRO
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/30/2021
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated at olympus (b)(4).(b)(4) checked the device and duplicated the reported phenomenon.As a result of the evaluation, the following was confirmed.-the front panel was flickering due to a malfunction of the filter turret.-the device has not been repaired in the last year.The exact cause of the reported event could not be conclusively determined.However, based on the following investigation results, the reported event may have been caused by the filter turret failure.The cause of the filter turret failure could not be determined.-(b)(4) confirmed that the front panel flashed during the evaluation and determined that it was caused by the filter turret failure.-omsc reviewed the manufacturing history and confirmed that there were no manufacturing abnormalities or corrections.Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed by the user that the front panel flashed after the procedure for inspection.The device was used in combination with the olympus otv-s7 series device and the procedure was completed without delay of more than 15 minutes.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA PRO XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13050036
MDR Text Key286251030
Report Number8010047-2021-16480
Device Sequence Number1
Product Code GCT
UDI-Device Identifier04953170228940
UDI-Public04953170228940
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S40PRO
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2021
Was the Report Sent to FDA? No
Date Manufacturer Received11/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2013
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-