Model Number 0070370 |
Device Problem
Component Misassembled (4004)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/30/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the holes in the front and back of the flat drain tube were misaligned.As per follow-up information received on 07-dec-2021, no other damage was found.
|
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
|
Manufacturer Narrative
|
The reported event was confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The product had caused the reported failure.Visual evaluation of the returned sample noted one unopened (with original packaging), hubless drain.A potential root cause for this failure could be ¿incorrect set-up".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "indications for use: wound drains are used to remove exudates from wound sites.Contraindications: do not use for chest drainage.Warning: do not bypass/inactivate anti-reflux valve." the actual/suspected device was evaluated.
|
|
Event Description
|
It was reported that the holes in the front and back of the flat drain tube were misaligned.As per follow-up information received on 07dec2021, no other damage was found.
|
|
Search Alerts/Recalls
|