MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD CORONARY DRUG-ELUTING STENT Back to Search Results
Lot Number 0025302693
Device Problems Break (1069); Positioning Failure (1158); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2021
Event Type  Injury  
Event Description
It was reported that difficulty stuck in lesion occurred resulting in surgery to remove a device fragment. A percutaneous coronary intervention was performed on a 90% stenosed, severely tortuous, and severely calcified lesion. A 2. 50 x 24mm synergy xd was advanced to the lesion, the balloon could not inflate, and the delivery system became stuck in the lesion. Attempts to recover the device using various other devices were made, but the synergy xd could not be recovered. While the synergy xd was being pushed and pulled, the shaft became torn. The procedure was discontinued, and the patient was sent to open heart surgery, where the device was finally removed. No further patient complications were reported in relation to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNERGY XD
Type of DeviceCORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13051872
MDR Text Key282554992
Report Number2134265-2021-16073
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date02/17/2022
Device Lot Number0025302693
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/21/2021 Patient Sequence Number: 1
-
-