DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-45
Device Problems Device Emits Odor (1425); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/26/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device (10/3233): testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to evaluate this unit.The fse found that the solenoid board had a smoked component which a damaged motor control board.The fse ordered the needed parts and returned to complete the repair on a subsequent day.The fse replaced the solenoid and motor control board with no success.The fse ordered the power management and backplane board.On a different date the fse replaced the power management board.A full calibration and functional check was performed per the factory calibration procedures.The unit was returned to the customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) shutdown unexpectedly and the screen went blank.In addition, the user stated that a high pitch squeal was heard and a burning odor was coming from the unit.A second cardiosave was retrieved and the patient therapy was transferred to that console.The therapy was resumed with minimal downtime.There was no harm or injury to the patient and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13052746
MDR Text Key282629678
Report Number2249723-2021-02915
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-45
Device Catalogue Number0998-00-0800-45
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/18/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age81 YR
Patient Weight69 KG
-
-