• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problems Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9735825, serial/lot #: (b)(4), ubd: unknown, udi#: unknown.A medtronic representative visited the site to evaluate the equipment.Hardware parts were replaced.(b)(4).The returned connector was analyzed.It was found the connector was bent.The rep was not able to plug in to test functionality.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a shunt placement procedure.It was reported that the lemo connector for the break out box was damaged and will no longer plug in.There was no reported impact to patient outcome and no reported delay to procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13054646
MDR Text Key286155494
Report Number1723170-2021-02934
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00763000359881
UDI-Public00763000359881
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
Patient Weight50 KG
-
-