Model Number CLV-S40PRO
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2021
Event Type  malfunction  
Manufacturer Narrative
The suspect device was returned to olympus and evaluated.The evaluation found that the spare lamp failed.The investigation is ongoing and the root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
A user facility reported to olympus that the spare lamp indicator was continuously lit.The problem, as reported to olympus, was identified during a procedure.The intended procedure was completed using the same device.There was no patient injury, associated with the problem, reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the phenomenon occurred due to the faulty spare lamp.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA PRO XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13057918
MDR Text Key286320204
Report Number8010047-2021-16587
Device Sequence Number1
Product Code GCT
UDI-Device Identifier04953170228940
UDI-Public04953170228940
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S40PRO
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse