Model Number G407376
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2021
Event Type  malfunction  
Event Description
During an atrial fibrillation procedure, after placing the sheath into the patient and prior to inserting any catheters or the transseptal needle, air was aspirated into a syringe that connected to the extension port of the sheath.Several aspirations were attempted, but the issue remained.The sheath set was replaced and the procedure was completed with no adverse consequences to the patient.
 
Manufacturer Narrative
Additional information: d9, g3, h2, h3, h6.One 8.5f swartz braided introducer sheath was received for evaluation.The sheath passed pressure and aspiration leak testing with no anomalies observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported leak remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL INTRODUCER, SL0¿, 8.5 F
Type of Device
TRANSSEPTAL CATHETER INTRODUCER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13061236
MDR Text Key283598971
Report Number3005334138-2021-00793
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG407376
Device Lot Number8023253
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial