Model Number OPT944
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The opt944 optiflow + adult nasal cannula is an interface used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt944 optiflow + adult nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information provided by the customer, and our knowledge of the product.Results: the healthcare facility reported that the tubing of an opt944 optiflow + adult nasal cannula broke during use.Conclusion: without the complaint device, we were unable to determine the cause of the reported event.However, the damage was likely caused by the tubing being pulled.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is rejected.The subject opt944 optiflow + adult nasal cannula would have met the required specifications at the time of production.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." "do not crush or stretch tube, to prevent loss of therapy." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Event Description
A healthcare facility in (b)(6) reported via a fisher and paykel healthcare (f&p) field representative that the tubing of an opt944 optiflow + adult nasal cannula broke during use.There were no reported patient consequences.
 
Manufacturer Narrative
(b)(4).The opt944 optiflow + adult nasal cannula is an interface used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt944 optiflow + adult nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information provided by the customer, and our knowledge of the product.Results: the healthcare facility reported that the tubing of an opt944 optiflow + adult nasal cannula broke during use.Conclusion: without the complaint device, we were unable to determine the cause of the reported event.However, the damage was likely caused by the tubing being pulled.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is rejected.The subject opt944 optiflow + adult nasal cannula would have met the required specifications at the time of production.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." "do not crush or stretch tube, to prevent loss of therapy." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Event Description
A healthcare facility in (b)(6) reported via a fisher and paykel healthcare (f&p) field representative that the tubing of an opt944 optiflow + adult nasal cannula broke during use.There were no reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key13064497
MDR Text Key285409321
Report Number9611451-2021-01487
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? No
Date Manufacturer Received11/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial