Catalog Number 393242
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  malfunction  
Event Description
It was reported that 61 bd venflon¿ pro safety shielded iv catheters' blister packs were discolored.The following information was provided by the initial reporter: "cannula packing is damaged.".
 
Manufacturer Narrative
Investigation summary: two photos were received by our quality team for evaluation.From the photos, a 22g unit pack and a 22g unit pack with top web product information was observed.The unit pack appears to be deformed; however, the details cannot be seen clearly from the photos.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.The manufacturing process was reviewed.There is no process in the manufacturing process that would cause the reported defect.There were no abnormalities during the production of the reported batch.Therefore, the reported defect could have occurred out of the manufacturing facility.The probable root cause for the deformed package could be due to the product being exposed to heat during handling (transportation, storage, loading and unloading, etc).This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13065264
MDR Text Key285405655
Report Number8041187-2021-01071
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number393242
Device Lot Number1203944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-