DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Manufacturer Narrative
During preventative maintenance (pm) conducted by a field service engineer (fse) the fse found this unit to have an electrical test fail code #42. Per service manual, the fse replaced the main pcb along with the software datasette. The fse performed a full pm per manufacture specifications, the unit passed all test and calibrations and is now ready for clinical use. A supplemental report will be submitted upon completion of our investigation. The initial reporter¿s name is shortened due to field character limit. The full name is (b)(6). The initial reporter named is a getinge employee who has different contact details from that of the event site. A contact person at the event site is: (b)(6).
 
Event Description
It was reported during preventative maintenance (pm) performed by a getinge field service engineer that the cs300 intra-aortic balloon pump (iabp) unit displayed electrical test fails code 42. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13081742
MDR Text Key287286302
Report Number2249723-2021-02957
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/22/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? No Answer Provided
Date Manufacturer Received12/03/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-