• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102208-500
Device Problem Imprecision (1307)
Patient Problem Hypoglycemia (1912)
Event Date 11/25/2021
Event Type  Injury  
Event Description
On (b)(6) 2021,senseonics was made aware of an incident where user experienced false hypoglycemia due to inaccuracies in sensor readings.
 
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Manufacturer Narrative
Based on the investigation, there was discrepancy with time of the reported event; the user indicated the event between 2-3am however the logs indicate the event at 6am.Upon review of the data between 2 am est and 7 am est, on (b)(6), it was observed there was inherent lag in the sensing technology of the sensor and that the sensor glucose eventually caught up to the calibration (capillary) entry within 3-4 sensor readings.Once the sensor performance recovered as the sensor use moved away from the insertion date, the system displayed better accuracy and there was good agreement between the sensor readings and fingerstick measurements following the event.User did not require medical intervention or treated herself as she wasn't symptomatic.As per dms, user is currently using the system with up to date information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key13085568
MDR Text Key282780399
Report Number3009862700-2021-00154
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491021441
UDI-Public00817491021441
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/25/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/12/2022
Device Model Number102208-500
Device Catalogue NumberFG-3300-01-001
Device Lot Number123501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/26/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
-
-