NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS10.0-50D235
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2021
Type of Reportable Event Malfunction
Event or Problem Description
Information was received that upon routine explant of the nail after intended treatment had been achieved, the nail was found to have corrosion.The nail was reported to have no other issues and the patient outcome was noted to be good.There was no patient harm or adverse impact reported.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional data: b5, g3, g6, h2, h3, h6, h10.Upon return, visual inspection of the nail revealed the distraction rod of the nail had visible evidence of discoloration at the junction of the nail and the distraction rod.The discoloration, also mentioned in the description as "corrosion" failure, is a previously known issue, which is being investigated under a capa.Visual inspection confirmed the reported discoloration failure mode.Following investigation, the nail has been quarantined.Review of the device history records indicated the precice-stryde nail was visually inspected and functionally tested prior to release.Functional testing was not applicable to the reported failure.In addition, the nail listed in this complaint was manufactured prior to the initiation of the capa.
 
Event or Problem Description
No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key13102042
Report Number3006179046-2021-00524
Device Sequence Number4696511
Product Code HSB
UDI-Device Identifier00887517951168
UDI-Public887517951168
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180503
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPS10.0-50D235
Device Lot Number9090903
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2021
Supplement Date Received by Manufacturer01/31/2022
Initial Report FDA Received Date12/28/2021
Supplement Report FDA Received Date02/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age17 YR
Patient SexMale
Patient Weight60 KG
-
-