• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNGYNA NOVO NORDISK GRAY; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNGYNA NOVO NORDISK GRAY; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNGYNA
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Insufficient Information (4580)
Event Date 12/14/2021
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Information received by medtronic indicated that the insulin pen had a leadscrew anomaly.Customer¿s blood glucose was unknown.No harm requiring medical intervention was reported.The insulin pen will be returned for analysis.
 
Manufacturer Narrative
Serial number: (b)(6).Software version: 3.8.5 color: grey battery life remaining: <11 months customer reports: unable to turn his pen past 4.5, states it starts to stick.Per visual inspection: no physical damage the inpen paired to the commercial app with small pairing dose.The screw is not bent.The leadscrew advances properly when dosing less that 3.0 unit, however, when dialing more than 4.0 units the leadscrew retracts and a high resistance its feel at the dial.Unable to perform functional test due to leadscrew anomaly.In conclusion: the customer complaint of unable to dial more than 4.5 units and leadscrew anomaly was confirmed.Pending further analysis that will be perform in san diego location.Per (b)(6), destructive testing showed that leadscrew anomaly was caused by a pattern wheel misalignment due to an encoder base bond failure.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNGYNA NOVO NORDISK GRAY
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key13102114
MDR Text Key285222935
Report Number3012822846-2021-00915
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000351
UDI-Public(01)000010862088000351(17)230521
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/21/2023
Device Model NumberMMT-105NNGYNA
Device Catalogue NumberMMT-105NNGYNA
Device Lot NumberB0195
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age60 YR
-
-