ABBOTT DIAGNOSTICS SCARBOROUGH, INC. BINAXNOW COVID-19 AG CARD; CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM.
|
Back to Search Results |
|
Catalog Number 195-000 |
Device Problem
False Negative Result (1225)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.A supplemental report will be provided after completion.Please also reference mfr.Report #s 1221359-2021-03435, 1221359-2021-02489, 1221359-2021-03769 thru 1221359-2021- 03774 and 1221359-2021-03775 thru 1221359-2021-03822.
|
|
Event Description
|
The customer reported 56 false negatives results with the binaxnow¿ covid-19 antigen card performed between (b)(6) 2021 and (b)(6) 2021.This mfr.Report addresses patient 55 of 56.The customer reported (1) false negative result with the binaxnow¿ covid-19 antigen card performed on (b)(6) 2021, using oropharyngeal swab sample.Pcr confirmation testing on oropharyngeal samples with amplitude generated positive results (ct values = 26.6).The customer reported that the patient was symptomatic.No additional patient information, including treatment and outcome, was provided.In addition, the customer has reported that they have sequestered 28587 individual kits of the binaxnow¿ covid-19 antigen card test for the mentioned lot number due to the false negatives the customer has received with the product.Since the customer confirmed the previously reported 7 false negative results are included in the 56 results mentioned an additional 49 reports will be submitted to address this complaint.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the investigation conclusion.Please see updates: additional information: the customer reported that there was no issue as of now as the instrument was replaced.Investigation: testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 131522 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 190-000 / lot 131522 and test base part number 190-430 / lot 131522.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 131522 showed that the complaint rate is (b)(4).In conclusion, the manufacturing batch record review (brr) revealed that the product met acceptance criteria for release, and review of complaints against the kit lot for the reported issue indicates that the product is performing according to the statements contained in the package insert and a product deficiency has not been identified.Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue; however, it could possibly be related to the specific patient samples.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the investigation conclusion.Please see updates: g3, g6, h2 and h6.Additional information: the customer reported that there was no issue as of now as the instrument was replaced.Investigation: testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 153364 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-160 / lot 153364 and test base part number 195-430h / lot 148134.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 153364 showed that the complaint rate is (b)(4).In conclusion, the manufacturing batch record review (brr) revealed that the product met acceptance criteria for release, and review of complaints against the kit lot for the reported issue indicates that the product is performing according to the statements contained in the package insert and a product deficiency has not been identified.Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue; however, it could possibly be related to issues including the self-test user performance, interpretation of the result, or the specific patient sample.
|
|
Manufacturer Narrative
|
This report is being submitted to correct the following fields of the intial report.Please see corrected fields: d1, d2, d3, d4, g4 and h8.
|
|