It was reported that the cs100 intra-aortic balloon pump (iabp) had a problem with the power supply, it was not powering on.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
Updated fields: b4, g4, g7, h2, h10, h11.Corrected fields: a1, b5, b6, b7, d11, e1 (event site address), h6 (component codes, health effect ¿ impact codes).The getinge field service engineer (fse) noted that the reported issue was reproduced and that after the power supply was replaced, the issue was solved.Subsequently, the fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted upon completion of our investigation.The full name of the event site in block e1 was shortened due to field character limit; the full name is (b)(6).An additional cusotmer contact name was provided: (b)(6).
|