• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC UNIVERSAL NAVIGATION EXPEDIUM SPINE SYSTEM QUICK CONNECT POLY DRIVER 5.5; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC UNIVERSAL NAVIGATION EXPEDIUM SPINE SYSTEM QUICK CONNECT POLY DRIVER 5.5; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 279734000N
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2021
Event Type  malfunction  
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021 during a spinal procedure, the tip of the quick connect poly driver broke during screw placement.The metal tip fragment was retrieved from the patient.There was no patient consequence.There was no surgical delay.There is no further information available.This report is for one (1) universal navigation expedium spine system quick connect poly driver 5.5.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Investigation summary: visual inspection: the 5.5 viper univ poly driver (part# 279734000n, lot# gm4222402 qty# 1) was returned and received at us cq.Upon visual inspection, it is observed that the distal tip of the driver shaft sub-assembly has broken off.The broken fragment was not returned.No issues were noted with the sleeve sub-assembly.Device failure/defect identified? yes.Dimensional inspection: dimensional inspection of the relevant feature of the received device was not performed at cq due to post manufacturing damage.Document/specification review: no design issues or discrepancies were found during this investigation.Complaint confirmed? yes.Investigation conclusion: the complaint is being confirmed for the 5.5 viper univ poly driver (part# 279734000n, lot# gm4222402 qty# 1) as the distal tip of the driver shaft was broken.A definitive root cause could not be identified for the reported issue from the available information.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based upon these results, no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: a review of the receiving inspection (ri) for 5.5 viper univ poly driver was conducted identifying that lot number gm4222402 was released in a two (2) batches.Batch1: lot was released on jan 20, 2015 with no discrepancies.Batch2: lot was released on mar 10, 2015 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device history review: the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL NAVIGATION EXPEDIUM SPINE SYSTEM QUICK CONNECT POLY DRIVER 5.5
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle 02400
SZ   02400
6103142063
MDR Report Key13114547
MDR Text Key287017944
Report Number1526439-2021-02673
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10705034388410
UDI-Public10705034388410
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K140927
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number279734000N
Device Catalogue Number279734000N
Device Lot NumberGM4222402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN SCREWS
Patient SexMale
-
-