Model Number M3535A
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2021
Event Type  malfunction  
Event Description
It was reported to philips that the battery does not charge.There was no patient involvement.
 
Manufacturer Narrative
It was reported to philips that the battery will not charge.The remote service engineer (rse) evaluated with the assistance of the customer and confirmed that the battery will need to be replaced.Upon conclusion of the evaluation, it was determined that this was a malfunction of the battery.The customer was given part information for a replacement battery and has ordered the part.The customer was informed of part availability and will ship once the part becomes available.The device remains at the customer site and no further evaluation is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
lois price
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key13114945
MDR Text Key282960360
Report Number3030677-2021-16710
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-