It was reported that the patient was not at target on the arctic sun device and showed an alert 113 (reduced water temperature control).
The patient temperature was 37.
3c, target temperature was 37.
0c, water temperature was 25.
6c, flow rate was 2.
0lpm, outlet monitor temperature (t1) was 25.
6c, outlet control temperature (t2) was 25.
6c, inlet temperature (t3) was 26.
4c, chiller temperature (t4) was 3.
7c, inlet pressure was -5.
8psi, circulation pump command was 100 percentage, mixing pump command was 100 percentage, system hours were 2994, pump hours were 2694.
Mss walked nurse through putting device into manual control with water temperature set to 10c.
Outlet monitor temperature (or) outlet control temperature remained at 26.
0c (or) 25.
9c, chiller temperature was 5.
2c, circulation pump command was 100 percentage, mixing pump command was 100 percentage.
Also confirmed that no one had added water to the device and advised the nurse how to disable manual control and to send to biomed tagged with alert 113.
|