• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY FUSION OMNI-TOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY FUSION OMNI-TOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Catalog Number FS-OMNI
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2021
Event Type  malfunction  
Event Description
In preparation for a procedure, the user selected a cook fusion omni-tome.It was initially reported that the physician opened the package at device preparation stage and checked the device functionality by injecting water to flush the cutting wire and found that the cutting wire was detached.There was no reportable information at this time.Our evaluation of the returned device on (b)(6) 2021 determined that the anchor had separated from the tip of the catheter.[subject of report] this occurred prior to patient contact; there was no impact to the patient.
 
Manufacturer Narrative
Pma/510(k) # k172288.Investigation evaluation: our laboratory evaluation of the product said to be involved determined the cutting wire securing component located near the distal end of the sphincterotome has disconnected from the catheter.The cutting wire is intact and remains securely attached to the sphincterotome at the proximal end.However, due to the catheter and securing component disconnection, the distal end of the cutting wire is no longer connected to the sphincterotome catheter at the distal end.The securing component has a longer section measuring 3.0 mm and a shorter section measuring 2.0 mm, therefore no part of the device is missing.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual prior to use testing conditions could not be duplicated in the laboratory setting.This limits our ability to conclusively determine a cause.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user: "do not over flex or bow tip beyond 90 degrees, as this may damage or cause cutting wire to break." other factors that can contribute to separation of the cutting wire securing component and the catheter include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all fusion omni-tomes are subjected to a visual inspection and functional test to ensure device integrity.The functional test includes bowing the sphincterotome to ensure the distal end responds to handle manipulation.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION OMNI-TOME
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key13118309
MDR Text Key285225814
Report Number1037905-2021-00730
Device Sequence Number1
Product Code KNS
UDI-Device Identifier10827002319032
UDI-Public(01)10827002319032(17)240611(10)W4480419
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFS-OMNI
Device Lot NumberW4480419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/06/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BOSTON SCIENTIFIC YELLOW ZEBRA WIRE GUIDE; ENDOSCOPE, UNKNOWN MAKE OR MODEL
-
-