• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752450
Device Problem Suction Failure (4039)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2021
Event Type  malfunction  
Event Description
A physician reported the aspiration failed during the ultrasound phase of a cataract procedure.The procedure was completed after replacing the product with another one.There was no patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.No adverse trends have been observed, associated with the reported product lot.The returned used/turbid sample was visually inspected, and no obvious defects were found.The non-invasive flow sensor (nifs) on the cassette housing was in good condition.A calibrated console representing the current software version was used to test the sample.The light-emitting diode (led) rings on the console turned green as the probe connectors were engaged to the console, indicating the proper communication between the probe and the console.The ball in the check valve of the drip chamber moved freely, per specification.The sample could prime and pass intraocular pressure (iop) calibration successfully.No anomalies were observed, during priming.No system message code was generated, during testing.The infusion pressure, irrigation, and aspiration rate were all measured at multiple set points throughout the console range and met specifications.Fluid flowed from the balanced salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed, after functional test had completed.The root cause of the customer's complaint could not be established.The returned sample functioned, per specifications.After investigation of this complaint.It has been determined, that this sample functioned per specifications.Therefore, no corrective action is required at this time.Quality assurance has reviewed this complaint.And will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13118456
MDR Text Key283082287
Report Number1644019-2021-00871
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657524501
UDI-Public00380657524501
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number8065752450
Device Lot Number2482582H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-