MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: a810; product type: software. If information is provided in the future, a supplemental report will be issued.
 
Event Description
H on (b)(6) 2021, information was received from an healthcare professional (hcp), via a manufacturer representative (rep), regarding a patient receiving morphine (unknown dose and concentration) via an implantable pump for spinal pain. It was reported that upon interrogation, the clinician programmer reported that the pump was at estimated replacement indicator (eri). However, the pump was implanted on (b)(6) 2021 with an initial eri estimate of 81 months. There were no known contributing factors. The hcp verified that they were in the correct patient's session versus a previous session. The correct patient name was showing on the tablet. The pump was re-interrogated after which pump reported an eri of 81 months again. Logs provided from (b)(6) 2021 state eri was "(b)(6)(1 months)".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13119692
MDR Text Key285340020
Report Number3004209178-2021-19169
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured10/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No