Model Number 19191
Device Problem Material Twisted/Bent (2981)
Patient Problem Hyperglycemia (1905)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
It was reported that the patient's blood glucose levels reached over 300 mg/dl while wearing the pod between 24 and 36 hours.When removed from the infusion site (abdomen), the pod's cannula was found bent.As treatment, a correction was delivered.
 
Manufacturer Narrative
According to the complainant the device will not be returned for investigation.We are unable to confirm the bent cannula or to determine if it could have contributed to the reported hyperglycemia.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including confirming the blue soft cannula is inspected for any damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key13120336
MDR Text Key283093561
Report Number3004464228-2021-23361
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385081120033
UDI-Public(01)20385081120033(11)210323(17)220923(10)L46152
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 12/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/23/2022
Device Model Number19191
Device Catalogue NumberZXP425
Device Lot NumberL46152
Was Device Available for Evaluation? No
Date Manufacturer Received12/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age17 YR
Patient SexFemale
-
-