The synergy xd mr ous 3.
50 x 20mm stent delivery system was returned for analysis.
A visual and microscopic examination of the stent found proximal stent damage, with stent struts lifted and pulled distally.
The undamaged stent od (outer diameter) was measured within maximum crimped stent profile measurement.
The balloon cones were reviewed, and no issues were noted.
The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.
A visual and microscopic examination of the bumper tip found distal tip damage.
A visual and tactile examination of the hypotube found a hypotube kink located 20.
3cm distal to the distal strain relief.
A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.
No other issues were identified during the product analysis.
|