U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM1272
Device Problems Incorrect Measurement (1383); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2021
Event Type  Death  
Event Description
Related manufacturer reference number: 2017865-2021-40498.It was reported that the patient has deceased.There is no known allegation from a health care professional that suggests the death was device related.No additional information was available.
 
Manufacturer Narrative
The reported event of incorrect measurement was not confirmed.Device image analysis indicated average current trends were normal and voltage was within normal range of operation.Analysis of device image showed that device was set to auto-capture and rate responsive mode was turned on.When device is set to auto settings the pacing may change based on requirements of the patient and may cause change in estimated longevity and battery measurement.Further analysis did not find any anomaly.Longevity assessment was performed, and device had appropriate longevity remaining.Correction: h6: health impact code should be "death" instead of " additional surgery".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13122138
MDR Text Key282995589
Report Number2017865-2021-40497
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509572
UDI-Public05414734509572
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model NumberPM1272
Device Catalogue NumberPM1272
Device Lot NumberA000105135
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2021
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age63 YR
Patient SexFemale
-
-