PHILIPS NORTH AMERICA LLC HEARTSTART MRX -EMS DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2021
Event Type  malfunction  
Event Description
It was reported to philips that the screen is broken and it does not charge.
 
Manufacturer Narrative
It was reported, that the device does not charge.There was reportedly no patient involvement.The customer requested, to return the device to the bench for evaluation and repair.The bench technician, evaluated the device.And determined, the physical damage to the handle assembly.The charging issue was not confirmed.Upon conclusion of the evaluation.It was determined, that this was a malfunction of the handle assembly, which was replaced.And the device was returned to full functionality.The device returned to the customer site.And no further evaluation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tinievangeline petty
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key13122533
MDR Text Key283091312
Report Number3030677-2021-16752
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838006652
UDI-Public00884838006652
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Device Catalogue NumberM3536A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-