• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Heart Failure/Congestive Heart Failure (4446); Mitral Valve Insufficiency/ Regurgitation (4451); Tricuspid Valve Insufficiency/ Regurgitation (4453)
Event Date 01/01/2021
Event Type  Injury  
Manufacturer Narrative
This information is based entirely on journal literature.Medtronic was made aware of this event through a search of literature publications.This event occurred outside the us.Patient information is limited due to confidentiality concerns.The event only occurred with one patient but specific details on the patient were not provided.The model listed in the report is a representative of the model family, as there is no specific model listed.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.A request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: heart failure caused by iatrogenic atrial septal defect after cryoballoon ablation for atrial fibrillation.Journal of cardiology cases.2021.24;303¿306.Doi.Org/10.1016/j.Jccase.2021.05.007.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A journal article was reviewed that contained information regarding cryoballoon ablation.The article reports a patient who underwent cryoballoon ablation.During the procedure of cryoballoon ablation, a single transseptal puncture was performed and a deflectable sheath and a second sheath were used with a circular mapping catheter and crossed the interatrial septum via the same puncture site.Approximately five months later, the patient was readmitted because of heart failure, recurrence of paroxysmal atrial fibrillation (af), and atrial flutter.Transesophageal echocardiogram (tee) showed a large iatrogenic atrial septal defect (iasd) and right-to-left shunt in the systolic and left-to-right shunt in the diastolic.Enlargement of the right ventricle was also observed, which caused annular dilatation as well as tethering of the tricuspid valve leaflets; hence, a moderate tricuspid regurgitation (tr) was observed.After optimal medical therapy for heart failure, elevation of pulmonary artery wedge pressure, mean pulmonary artery pressure, and pulmonary to systemic blood flow ratio was revealed.A second radiofrequency catheter ablation (rfca) was performed seven months after the first session and tee showed persistent iasd with a bidirectional shunt.Eleven months after the first ablation, the patient was again readmitted because of heart failure, recurrence of paroxysmal af, and atrial tachycardia.Persistent iasd with bidi rectional shunt remained.Therefore, a percutaneous transcatheter closer was performed because heart failure could not be controlled by medication.After the additional third catheter ablation for re-isolation of the right pulmonary vein, defragmentation of the left atrium, and superior vena cava isolation, the iasd was closed.Heart failure was controlled with a decrement of right atrial and ventricular size, and improvement of tr (from severe to mild).Mitral regurgitation also improved from moderate to mild, and there was no shunt at the interatrial septum.The status/disposition of the sheath and catheter is unknown.No further patient complications have been reported as a result of this event.Further follow up did not yet yield any additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13156011
MDR Text Key285698312
Report Number3002648230-2022-00001
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age83 YR
Patient SexMale
Patient RaceAsian
-
-