B4; g1, 3, 6; h2, 3, 6 h3: the investigation by ge healthcare has been completed.The acoustic performance test was performed on the system and concluded that the testing meets the iec 60601-2-33 requirements and the osha levels are within the specification for this system configuration.The incident appears to be the result of human medical condition(s).The patient was provided specified hearing protection during the scan.Human condition may cause sensitivity to acoustic levels that occur during normal clinical scanning.No system issue was found.No corrections are required as the system was operating within specification.
|