• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRATOME XL; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRATOME XL; SNARE, FLEXIBLE Back to Search Results
Model Number M00535900
Device Problems Break (1069); Failure to Fold (1255)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2021
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an ultratome xl was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, the cutting wire did not bow.The procedure was completed another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed a reportable event based on the investigation results: the cutting wire was broken.
 
Manufacturer Narrative
(b)(4).The returned ultratome xl was analyzed, and a visual evaluation noted that the cutting wire was broken approximately at 18mm from distal pierced hole, and blackened, consistent with the findings when the device was observed under magnification.The working length was free from obvious kinks and bends.Functional evaluation was not performed due to the device condition.No other problems with the device were noted.The product analysis revealed that the cutting wire was broken, and blackened.The cutting wire being blackened indicates that the device was energized.These conditions could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during procedure.Energizing the device prior to performing sphincterotomy can also compromise the cutting wire's integrity and can cause a premature cutting wire fatigue, leading to a break.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATOME XL
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13168114
MDR Text Key283245870
Report Number3005099803-2021-08102
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729103257
UDI-Public08714729103257
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K930022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model NumberM00535900
Device Catalogue Number3590
Device Lot Number0027237374
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight67 KG
-
-